FAQs

50/600mL bioreactor / test tube

Where can I find appropriate operating conditions for TubeSpin Bioreactors?

We provide a summary of references from literature and Dr. Shaker's recommendations.

Dr. Shaker`s Recommendation


Mammalian Cell cultureInsect CellsPlant Cell culture
TubeSpin 50CHOHEK-293SF-9N. tabacum BY-2
Shaking Speed [rpm]180 - 250180 - 250180 - 250180 - 250
Throw [mm]50505050
Working Volume [mL]15 - 3015 - 305 - 105 - 10





TubeSpin 600CHOHEK-293SF-9N. tabacum BY-2
Shaking Speed [rpm]180 - 250180 - 250180 - 250180 - 250
Throw [mm]50505050
Working Volume [mL]300300200150

Methods from Literature

TubeSpin 50CHOCHOCHOCHOSF-9
Shaking Speed [rpm]180-300160-200140200160-200
Throw [mm]50n.s.5025n.s.
Working Volume [mL]14-3410-355-10510-35
ReferenceStrnad (2010), Biotechnol. Prog.: Vol. 26 (3); pp 653-663Xie (2011), Cytotechnology: Vol. 63 (4); pp 345-350Stettler (2007), Biotechnol. Prog.: Vol. 23 (6); pp 1340 -1346De Jesus (2004), Biochem. Eng. Journal: Vol 17, No. 3; pp 217 - 223Shen (2011), BMC Proc.: 5 (Suppl 8); P37






Tube Spin 600CHOCrypthecodinium cohnii (marine microalgae)


Shaking Speed [rpm]120-220230


Throw [mm]5025


Working Volume [mL]300100


ReferenceMonteil (2013), Biochem. Eng. Journal; Vol. 76, pp 6–12Hillig (2014), Adv Biochem Eng Biotechnol; Vol. 138: pp 179-206